Effects of extraction and lens voltages in extracting H− ions from a magnetized sheet plasma using an E×B probe
Abstract
This paper discusses the effects of Vext and Vlens parameters of an E×B probe in the extraction of H− in a magnetized sheet plasma. Variation of Vext shows that Ipeak is maximum at Vext = 190V and FWHM is narrower at higher Vext. Whereas the variation of Vlens reveals that Ipeak and FWHM are not much affected by negative Vlens but are decreased by increasing positive Vlens.
Downloads
Published
2002-10-23
Issue
Section
Condensed Matter and Plasma Physics
How to Cite
[1]
“Effects of extraction and lens voltages in extracting H− ions from a magnetized sheet plasma using an E×B probe”, Proc. SPP, vol. 20, no. 1, pp. SPP–2002, Oct. 2002, Accessed: Apr. 23, 2026. [Online]. Available: https://proceedings.spp-online.org/article/view/SPP-2002-2D-05