Confronting the issue of phase drift and tuning in AC susceptibility measurements using a mutual inductance bridge

Authors

  • Roland V. Sarmago ⋅ PH National Institute of Physics, University of the Philippines Diliman

Abstract

To elucidate the origin and manifestation of phase drift, to minimize the effect, and to interpret current data on AC susceptibility, a detailed phasor analysis of a mutual inductance bridge (MIB) is presented. Understanding responses of the bridge with frequency and field with temperature will allow one to implement a protocol for tuning the MIB properly.

Downloads

Published

2002-10-23

How to Cite

[1]
“Confronting the issue of phase drift and tuning in AC susceptibility measurements using a mutual inductance bridge”, Proc. SPP, vol. 20, no. 1, pp. SPP–2002, Oct. 2002, Accessed: Apr. 23, 2026. [Online]. Available: https://proceedings.spp-online.org/article/view/SPP-2002-2C-03