Effect of varying Cs-137 beam on the response of a radiation monitoring instrument (survey meter)
Abstract
The effect of varying Cs-137 beams, by utilizing a lead attenuator and a collimator, on a survey meter’s (SM) response was assessed in this study. The procedures on the calibration method of SM in the Philippine Nuclear Research Institute–Secondary Standards Dosimetry Laboratory (PNRI–SSDL) were presented.
Their resulting measurements in terms of the ambient dose equivalent H*(10) were shown. Results showed that at source-detector distances with lower H*(10), bare Cs-137 beam produced better response for the SM. However, it is important to note that the use of lead attenuator and collimator are still more practical when it comes to procedures in the PNRI-SSDL. This decreases the source-detector distance needed for the calibration. Distance may have also played a role as to why an instrument response is closer to the reference value.
Downloads
Published
Issue
Section
License
By submitting their manuscript to the Samahang Pisika ng Pilipinas (SPP) for consideration, the Authors warrant that their work is original, does not infringe on existing copyrights, and is not under active consideration for publication elsewhere.
Upon acceptance of their manuscript, the Authors further agree to grant SPP the non-exclusive, worldwide, and royalty-free rights to record, edit, copy, reproduce, publish, distribute, and use all or part of the manuscript for any purpose, in any media now existing or developed in the future, either individually or as part of a collection.