Effect of camera digitization in filter-based fluorescence spectral microscopy

Authors

  • Luis Buño National Institute of Physics, University of the Philippines Diliman
  • Caesar A. Saloma National Institute of Physics, University of the Philippines Diliman

Abstract

Image digitization of optical images was predicted to limit the accuracy of spectral estimates obtained in filter-based fluorescence spectral microscopy. We investigate experimentally the issue using two 3-CCD cameras - an 8-bit Hitachi HV-C20 and a 12-bit Hamamatsu ORCA C7780-20. Spectral measurement is achieved using a weighted superposition of basis spectra that are derived from a spectral library. Multi-channel detection using three independent camera channels and color filters allows the calculation of coefficients for the basis spectra. The emission spectra of yellow-green and orange fluorescent microspheres were measured. More accurate estimates are derived using the 12-bit camera compared to the 8-bit CCD camera.

Downloads

Issue

Article ID

SPP-2010-PA-22

Section

Poster Session PA

Published

2010-10-25

How to Cite

[1]
L Buño and CA Saloma, Effect of camera digitization in filter-based fluorescence spectral microscopy, Proceedings of the Samahang Pisika ng Pilipinas 28, SPP-2010-PA-22 (2010). URL: https://proceedings.spp-online.org/article/view/SPP-2010-PA-22.