Assessment of leakage and scattered radiation of C-arm fluoroscopy units

Authors

  • Bianka Mandac ⋅ PH Department of Physical Sciences and Mathematics, University of the Philippines Manila
  • Sarah Johnson ⋅ PH Department of Physical Sciences and Mathematics, University of the Philippines Manila
  • Alvin Baclig ⋅ PH Department of Physical Sciences and Mathematics, University of the Philippines Manila
  • Isabel Elona ⋅ PH Center for Device Regulation, Radiation Health, and Research, Department of Health

Abstract

C-arm fluoroscopy has been an integral part of both interventional and cardiac procedures. But despite the benefits a C-arm fluoroscopy unit can provide, there is a substantial risk associated with its use for the workers with regards to the leakage and scattered radiation from the C-arm. Using a survey detector, the researcher measured the leakage and scattered radiation of two C-arm units at different distances and angles. It was observed that increasing the distance from the radiation source decreases the radiation exposure and that radiation exposure is dependent on the positioning of the C-arm.

Downloads

Issue

Article ID

SPP-2015-4C-05

Section

Complex Systems, Atmospheric Physics, Biophysics, and Medical Physics

Published

2015-06-03